Education

Overview Of Temporary Pacemakers

Pinterest LinkedIn Tumblr

The purpose of temporary pacemakers is to sense intrinsic electrical activity and pace the myocardium from the inside (ie, transvenous wires) or outside (epicardial wires). Following cardiac surgery, surgeons often place bipolar leads on the right ventricle (RV) for ventricular sensing/pacing +/- right atrial (RA) leads for synchronized atrioventricular pacing. A current is delivered between the two bipolar leads and “captured” by the myocardium resulting in depolarization.

There are three major settings for pacers:

  • Rate: the number of impulses/minute
  • Output (mA): the current needed for the myocardium to “capture” the paced signal. In other words, how energy much are you jolting the heart with. Atrial output and ventricular output range from 0.1 – 20 mA and 0.1 – 25 mA, respectively.
  • Sensitivity (mV): the point at which the pacing device interprets a native myocardial signal as a pulse. If sensitivity is too low, then any – little – signal (including breathing, patient movement, etc.) will be interpreted as a native beat (oversensing). The pacer might see no need to fire which could result in bradycardia/asystole. If the sensitivity is too high, then native activity might be missed all together (undersensing) causing the pacer to potentially fire on a native heart beat (ie, R-on-T phenomenon).

To “test” a pacer, I select a mode (typically DDD for A-V pacing wires and VVI for only V pacing). I’ll start with the default output and sensitivity settings and increase the pacer rate to something above the patient’s native heart rate. Then I’ll watch the continuous EKG to ensure that the signal is indeed captured at the preset rate while dialing down the output.

If a pacer loses its ability to capture, one must consider possible changes in sensitivity thresholds, electrolyte changes, damaged lead(s), battery depletion, or faulty cables. If I seem to have competition between the pacer and the patient’s native heart rate, I’ll adjust the sensitivity to avoid the aforementioned scenarios of over/undersensing.

Drop me a comment below with questions! Also, be sure to check out my more thorough post on pacemaker nomenclature and whether or not a magnet is needed.

Write A Comment

Pin It