Measuring And Interpreting Intracranial Pressure (ICP) Waveforms

Managing elevated intracranial pressure (ICP) is of paramount importance in neurocritical care. Placing bedside ICP monitors are contraindicated if the patient is coagulopathic, has a brain abscess, or has an active scalp infection. Otherwise, the procedure is fairly straightforward and usually boils down to two choices: a Codman ICP monitor (aka, “bolt“) or an external ventricular drain (EVD).

CT head (coronal view) showing EVD catheter in the left lateral ventricle

The bolt is a catheter with a strain gauge tip which sits in the brain tissue (parenchyma). Strain gauge basically means a change in force applied to the tip (ie, changes in ICP) will concurrently change electrical resistance within the catheter. FYI,  it’s called a “bolt” because a bolt literally affixes the catheter at the cranial suture line.

In comparison, an EVD sits in the third ventricle or lateral ventricles and can be used for measuring ICP as well as therapeutic purposes when draining cerebrospinal fluid (CSF). EVDs are more invasive and often times more difficult to place if the ventricular network is displaced or collapsed.

Both the bolt and EVD can transduce a waveform which has some similarities to an arterial line (A-line) waveform. In the normal waveform above, P1 is the “percussion wave” and represents arterial pulsation just like the initial upstroke on an A-line. P2 is the “tidal wave” (cool name, 😎 ) and represents intracranial compliance. Compliance is defined by change in volume / change in pressure, so a poorly compliant brain will accept any additional volume (blood, CSF, brain swelling) by rapidly increasing ICP. P3 is the dicrotic wave, represents a reflected pressure wave due to aortic valve closure, and is analogous to the dicrotic notch on an A-line.

Another important aspect to understand is that the waveform’s overall amplitude will depend on CSF volume. If you’ve just drained a large amount of CSF with an EVD, the new waveform amplitude will decrease in comparison to before drainage. The opposite is also true – accumulation of CSF will augment the waveform amplitude.

Drop me a line with any questions below! 🙂